<ArticleTitle>Numerical Study of Microbubble Dynamics Subjected to Ultrasound and Its Effect on Thermal Ablation of Biological Tissue</ArticleTitle>
<VernacularTitle>Numerical Study of Microbubble Dynamics Subjected to Ultrasound and Its Effect on Thermal Ablation of Biological Tissue</VernacularTitle>
			<FirstPage>2561</FirstPage>
			<LastPage>2582</LastPage>
			<ELocationID EIdType="pii">5044</ELocationID>
			
<ELocationID EIdType="doi">10.22060/mej.2023.21626.7480</ELocationID>
			
			<Language>FA</Language>
<AuthorList>
<Author>
					<FirstName>Ghassem</FirstName>
					<LastName>Heidarinejad</LastName>
<Affiliation>Faculty of Mechanical Engineering, Tarbiat Modares University, Tehran, Iran</Affiliation>

</Author>
<Author>
					<FirstName>Afsaneh</FirstName>
					<LastName>Mojra</LastName>
<Affiliation>Faculty of Mechanical Engineering, K. N. Toosi University of Technology, Tehran, Iran</Affiliation>
<Identifier Source="ORCID">0000-0001-8790-7016</Identifier>

</Author>
<Author>
					<FirstName>Hossein</FirstName>
					<LastName>Azizi Sormoli</LastName>
<Affiliation>Faculty of Mechanical Engineering, Tarbiat Modares University, Tehran, Iran</Affiliation>

</Author>
</AuthorList>
				<PublicationType>Journal Article</PublicationType>
			<History>
				<PubDate PubStatus="received">
					<Year>2022</Year>
					<Month>07</Month>
					<Day>25</Day>
				</PubDate>
			</History>
		<Abstract>High-intensity focused ultrasound is a non-invasive method and provides many therapeutic applications for physicians. One of the ways to increase the efficiency of High-intensity focused ultrasound is using a Levovist contrast agent, which consists of microbubbles. In the present study, we calculate the pressure field due to the High-intensity focused ultrasound using the Helmholtz equation for linear ultrasonic wave propagation. Using the Keller-Miksis equation, we calculate the thermal effects caused by microbubble injection after determining the acoustic pressure. The Pennes bioheat transfer equation is used for studying the tissue temperature distribution. The simulation results show that in the presence of a microbubble under the influence of a High-intensity focused ultrasound pressure field, increasing the applied frequency and power increases the value of heat sources caused by the microbubble oscillation. An increase in the temperature of biological tissue can be observed after the injection of microbubbles. Within the pressure range of 2.54 MPa, the tissue temperature at the focal point, for the case where the microbubble with the initial radius of 50 μm is injected, increases by 8.28 ℃. Meanwhile, if a microbubble with an initial radius of 50 micrometers is injected, there is a further increase in the tissue temperature by 57.72%. In the absence of microbubbles, the corresponding temperature rise is only 5.42 ℃ for the same operating conditions. Finally, the Arrhenius model shows that the microbubbles with different initial radii increase the ablated tissue volume by about 38%.</Abstract>
			<OtherAbstract Language="FA">High-intensity focused ultrasound is a non-invasive method and provides many therapeutic applications for physicians. One of the ways to increase the efficiency of High-intensity focused ultrasound is using a Levovist contrast agent, which consists of microbubbles. In the present study, we calculate the pressure field due to the High-intensity focused ultrasound using the Helmholtz equation for linear ultrasonic wave propagation. Using the Keller-Miksis equation, we calculate the thermal effects caused by microbubble injection after determining the acoustic pressure. The Pennes bioheat transfer equation is used for studying the tissue temperature distribution. The simulation results show that in the presence of a microbubble under the influence of a High-intensity focused ultrasound pressure field, increasing the applied frequency and power increases the value of heat sources caused by the microbubble oscillation. An increase in the temperature of biological tissue can be observed after the injection of microbubbles. Within the pressure range of 2.54 MPa, the tissue temperature at the focal point, for the case where the microbubble with the initial radius of 50 μm is injected, increases by 8.28 ℃. Meanwhile, if a microbubble with an initial radius of 50 micrometers is injected, there is a further increase in the tissue temperature by 57.72%. In the absence of microbubbles, the corresponding temperature rise is only 5.42 ℃ for the same operating conditions. Finally, the Arrhenius model shows that the microbubbles with different initial radii increase the ablated tissue volume by about 38%.</OtherAbstract>
		<ObjectList>
			<Object Type="keyword">
			<Param Name="value">Focused ultrasound</Param>
			</Object>
			<Object Type="keyword">
			<Param Name="value">biological tissue</Param>
			</Object>
			<Object Type="keyword">
			<Param Name="value">Microbubble dynamics</Param>
			</Object>
			<Object Type="keyword">
			<Param Name="value">Thermal ablation</Param>
			</Object>
			<Object Type="keyword">
			<Param Name="value">numerical simulation</Param>
			</Object>
		</ObjectList>
<ArchiveCopySource DocType="pdf">https://mej.aut.ac.ir/article_5044_996740de914ced0902e686373e319391.pdf</ArchiveCopySource>
</Article>
</ArticleSet>